Designed as a direct replacement for the
popular model UCS20, the UCS30 nuclear spectrometer offers many enhancements
and additional features for increased flexibility. USB connectivity provides
plug and play operation with most of computers running Windows 2000 or XP.
Three models are available with 1024, 2048 and 4096 channels of conversion
gain.
The UCS30 ADVANCED Spectrometer System is capable for
conducting a wide selection of spectroscopy experiments. This external unit
interfaces through a USB port and features a 1024 channel MCA with internal
preamplifier, computer controlled amplifier, high voltage (0-2048V), upper and
lower level discriminators and Multichannel Scaling for half-life and decay
studies. Also included is a Mossbauer interface and preamplifier power output
(±12 V) for connecting detectors with external preamplifiers such as
proportional counters and CdZnTe probes.
The complete system includes a NaI(Tl) scintillation
detector with lead shield and multi-position sample stand, a set of 8 gamma
emitting radioisotope sources with an “unknown”, MCA software with ISOMATCH
peak identification and a lab manual of experiments in gamma spectroscopy on
CD.
System Components Include:
- UCS30-1K
Universal Computer Spectrometer with Windows 2000/XP Software.
- SPA 38
Complete scintillation probe with SDA38 detector, stand, base and cables.
- RSS 8
Set of 8 gamma sources including "unknown".
- LM 5
Experimental Gamma Ray Spectroscopy manual on CD.
- USB 6
USB cable.
The internal preamplifier allows direct
connection to scintillation detectors. The bipolar amplifier includes polarity
selection and multiple shaping times when using different types of detectors
with external preamplifiers. Upper and lower level discriminators may be set
for rejection of unwanted signals. The high voltage supply is rated to +2048V
for use with scintillation or proportional detectors. Optional polarity
selection is available for use with detectors requiring negative bias. A
preamplifier power connector (±12V) is
included for use with our model PA-1 external preamplifier. A 10µsec fixed
conversion ADC allows high throughput with deadtime correction. Input
selection for preamplifier, amplifier or ADC direct is included by computer
control.
In addition to the normal pulse height
analysis mode the UCS30 may be operated as a Multichannel Scaler for time
correlated studies such as half-life decay. A special Mossbauer mode is
included with MSB output for synchronization of the drive circuit. X-ray
mapping may be performed with the optional ROI output.
For ease of setup an auto-calibrate mode is
provided that automatically sets the HV and amplifier gain to 1 keV/ channel
using a Cs-137 source. Once the system is calibrated all parameters may be
saved as setup files for quick recall of specific measurements. Data may be
saved as binary files for reloading into memory or as spreadsheet compatible
files for data processing.
A background buffer is included for
storing a spectrum when background subtraction, factional spectrum stripping
or overlay is required. Multiple ROIs may be set with calculation of gross
and net area. Presets include Real Time, Live Time and ROI Integral.
Isomatch includes a text editable library
of common isotopes. When loaded, markers are placed on the display at the
positions of the isotope energies to aid in identification.
Multiple UCS30’s may be connected to a
single computer providing USB ports are available. The internal power supply
is UL, CE and CSA listed for operation worldwide. Warranty is one year parts
and labor, return to factory.
Specifications.
Hardware.
· Physical: Bench top enclosure with
USB 1.1 interface includes preamplifier, amplifier, detector high voltage,
up to 4096-channels depending on model, Multichannel Analyzer with data
memory, LLD and ULD. Pulse height analysis and Multichannel Scaling modes.
Fully compatible with many scintillation detectors and commercial tube
bases.
· Data Input: BNC connector
multi-function input. On-board relays select internal preamplifier,
shaping amplifier or direct to ADC when using external signal processing.
· Gate Input: BNC connector.
Normally open gate. TTL low closes gate to ADC.
· HV Output: MHV connectors.
Positive output, (switched positive or negative outputs with HV option).
(SHV connectors optional), 0-2048 volts, 1mA maximum current at 1000
volts. High voltage is fully regulated and computer controlled in 2 volt
increments.
· MCS Input: BNC connector bypasses
internal preamp/amplifier. Accepts positive TTL signal >150nsec., 5MHz
maximum rate.
· ROI Output: BNC connector.
Generates pulse each time an event occurs in set region of interest.
Requires ROI option.
· MSB Output: BNC connector. TTL
output of dwell counter most-significant-bit for synchronizing Mossbauer
drive.
· USB: USB type “A” connector. USB
1.1 standard.
· Preamplifier: Charge sensitive
preamplifier for use with scintillation detectors and PMTs.
· Amplifier: Bipolar shaping
amplifier with selectable peaking times of 1, 2, 4 and 8µsec. Computer
controlled coarse and fine gain from x1 to x160. Input polarity selection.
· External Preamp Power: DB-9F
connector supplies ±12 volts.
· ADC: Fixed conversion 10usec.
Computer selected conversion gain from 256 to 4096 channels depending on
model. Direct input accepts pulse peaking times of 0.5µsec. to 10µsec.
Includes dead-time correction when used in Live-time mode.
· LLD & ULD: Independently computer
controlled discriminators over entire input range. Operates prior to ADC
for reduced system dead time.
· Modes: MCA for pulse height
analysis, MCS for half-life decay or other time related studies and
Mossbauer mode.
· Timers: Real-time or Live-time
operation selectable in 1-second increments for PHA, or dwell times from
10msec. to 60sec. per channel in MCS mode. In Mossbauer mode, 100usec. to
500usec.
· Data Memory: On-board dynamic RAM,
up to 4096 channels x 3 Bytes for data, plus region-of-interest flag.
· Deadtime: System dead time is
computed and displayed on screen during acquisition.
· Power: AC line, 100-250VAC,
50/60Hz, auto sensing internal power supply with Universal Power Cord
input.
Software.
· System: Operates under Windows
2000/XP.
· Display: Color or monochrome
monitor, Vertical scale adjusts from 32 to 16M and LOG display. Horizontal
4096 channels depending on model with expansion down to 128 channels.
· Time Mode: Preset live-time or
preset real-time selection. Both times are recorded and displayed.
· Count Mode: Preset count mode
integral of region-of-Interest.
· ROI'S: Multiple Regions-of
Interest using color coding.
· Integral: When cursor is in ROI,
computes gross area, net area with end point averaging, centroid and FWHM.
· Energy Cal: 2-point linear or
3-point quadratic converts cursor position reading directly to energy
units. (Time units in MCS mode.)
· Autocalibrate: Use with Cs-137
source to automatically calibrate system for 1 keV/channel.
· Bkgd Buffer: Used to store
background spectrum for time normalized background subtraction.
· Subtract: Subtracts
channel-by-channel time normalized data stored in Background Buffer from
display data.
· Smooth: 3-point smoothing of
displayed data. Does not modify data in acquisition memory.
· Control: Software control of Input
Selector, High Voltage and Polarity (with HV option), Amplifier Input
Polarity, Peaking Time, Gain, Lower and Upper level discriminators, and
ADC Conversion Gain.
· File: Save or load data file and
header information in binary format or as CSV/TSV files. Save or load
setup files for specific configurations.
· Clipboard: Clips data and header
information, graphic screens, ROI data containing, centroid, FWHM, gross
and net integrals, and header information in spreadsheet compatible format
for transferring into other applications.
· ISOMATCH: Isotope library text
file with peak markers and labeling for overlaying on spectrum. Provides
quick isotope identification. Libraries may be edited and extended using
word processors or text editor programs.
Ordering Information.
- ADV1K Universal Computer
Spectrometer System 1024 channel for Windows 2000/XP
- ADV2K Universal Computer
Spectrometer System 2048 channel for Windows 2000/XP
- ADV4K Universal Computer
Spectrometer System 4096 channel for Windows 2000/XP
- UCS30-HV Positive and negative high voltage
option
- UCS30-ROI Region of Interest pulse
output option
Accessories
- LS325 Environmental lead shield for 2" or 3" detectors.
- SD51 2x2 in. NaI(Tl) with integral photomultiplier.
- TS51-2 Tube base stand, 14-pin for 2" or 3" detectors,
complete with cables.
- PA1 External preamplifier for proportional
counters.
- SP78W-2 Well detector complete with stand, lead shield and
cables.
- SPA 38 Complete detector assembly with 38x38mm detector,
stand, shield and cables.
|